Soredex Cranex D Ceph
 
 

X-ray generator

TUBE

- OPX/105, or equivalent

FOCAL SPOT

- 0.5 mm IEC 336

TARGET ANGLE

- 5º

TARGET MATERIAL

- Tungsten

OPERATING TUBE POTENTIAL

- Panoramic imaging 57 - 85 kV (±4 kV)

- Cephalometric imaging 60-85 kV (±4 kV)

OPERATING TUBE CURRENT

- 10 mA (±1 mA) at 0.5 FS

MAXIMUM TUBE CURRENT

- 11 mA

MAXIMUM OUTPUT POWER

- 945 W nominal

FILTRATION

- minimum filtration 2.7 mm Al

BEAM QUALITY

- HVL over 2.5 mm Al @ 85 kV

OUTER SHELL TEMPERATURE

- +50ºC (122ºF) maximum

DUTY CYCLE

- controlled by the software of the unit

Power requirements

INPUT VOLTAGE

- 230 or 115 VAC (±10%), 50/60 Hz, single phase, grounded socket

MAXIMUM LINE CURRENT

- 7 A (@85 kV/10mA, 230 VAC mains)

MAXIMUM LINE RESISTANCE

- 1 ohm

MAXIMUM LINE FUSING

- 10 A /20A slow @ 230/115 VAC (main fuse 8A/16A slow in the device)

LINE SAFETY SWITCH (when required)

- Approved type, min. 10 A 250 VAC

EARTH LEAKAGE CIRCUIT BREAKER (when required)

- Approved type, min. 16 A 250 VAC, breaker activation leakage current

accordance with local regulations.

Mechanical parameters

PANORAMIC

- Source to Image layer Distance (SID) 520 mm (±10 mm)

- Magnification factor 1.34

CEPHALOMETRIC

- Source to Image layer Distance (SID) 1721 mm ±20 mm

- Source to Object Distance (SOD) 1500mm

-Magnification factor 1.15

WEIGHT

- Panoramic unit 120 kg

- Panoramic/cepahlometric unit 165 kg

DIMENSIONS

- Panoramic unit (H x W x D) 2320 x 1200 x 1000 mm

- Panoramic/cephalometric unit (H x W x D) 2320 x 1200 x 1900 mm

VERTICAL HEIGHT OF CHIN REST

- 950 - 1750 mm (+- 10 mm)

Digital image receptor

Only the CCD sensors specifically designed for Cranex D unit can be used.

PIXEL SIZE

- 96 micrometres

Timer

PANORAMIC EXPOSURE TIMES

- Normal 17.6 s (±15%)

- Child 16 s (±15%)

- Partial 1.9 s - 3 s - 9.2 s - 3 s - 1.9 s

Can be freely selected and combined, overlapping approx. 0.3 s.

- TMJ 3.3 + 3.3 s (±15%)

Max 240 mAs

CEPHALOMETRIC EXPOSURE TIMES

- 8 - 20 s scanning times, 5 steps according to R’10 series (ISO)

BACK-UP TIMER

- 23.5 s (±1.5s)

Leakage technique factors

PANORAMIC

- 85 kV, 2400 mAs/h (85 kV, 10 mA, duty cycle 1:15)

CEPHALOMETRIC

- 85 kV, 1800 mAs/h (85 kV, 10 mA, duty cycle 1:20)

Measurement bases

kV and mA values can be verified with a specified digital multimeter according to

separate measurement instructions. The exposure times can be measured as the

duration of radiation in the primary radiation beam.

Exposed field size in cephalometry

- 22 x 26 cm for lateral projections

- 22 x 22 cm for PA and AP projections

- Automatic filtration of soft tissues for lateral projections controlled by software.

Operating ambient conditions

- Operating temperature 10 - 40ºC

- Relative humidity 0 - 85 RH%

Storage ambient conditions

- Storage temperature 0 - 40ºC

- Relative humidity 0 - 85 RH%

Minimum computer requirements

The values in (brackets) are recommended values.

OPERATING SYSTEM

- Windows XP Professional / Home / SP1 or SP2

- Windows 2000 Professional / SP4

CPU

- Pentium 4 or Athlon XP or equivalent (1.5 GHz or better recommended)

RAM

- 256 MB (512 MB recommended)

HDD

- 20 GB (single user)

VIDEO RAM

- 16 MB (or more)

NETWORK CONNECTION

- 10/100 Mbit/s Ethernet NIC

DISPLAY

- 1280 x 1024 x 24-bit Tru Color, 85Hz display

(19" CRT or 17” TFT LCD recommended)

COM 1 and COM 2 available (free)

Connection to the PC must meet EN60601-1 requirements.

The use of ACCESSORY equipment not complying with the equivalent safety requirements

of this equipment may lead to a reduced level of safety of the resulting

system. Consideration relating to the choice shall include:

- use of the accessory in the PATIENT VICINITY

- evidence that the safety certification of the ACCESSORY has been performed in

accordance to the appropriate IEC 601-1 and/or IEC 601-1-1 harmonized national

standard

- only RS 232C interface cable and fibre cable, provided by the manufacturer, shall

be used.

 
 
Copyright 2003 Smitech (Asia) Pte Ltd. All rights reserved.